DT Series - Características
Differential Transformer Type Extensometers
Extensometer Class 1 Compliant
This instrument is compliant with Class 1 in ISO 9513 and JIS B 7741
(±3 μm or ±1 % of the specified value, whichever is larger).
It is also compliant with ASTM E83 Class C.
Compliant with Strain Rate Control Test Methods
Compliance with Extensometer Class 1 and Strain Control
Please reference the table below when using DT series differential transformer type extensometers for Class 1 compliance and when implementing strain rate control.
UH-X/FX/ AG-X/ AGS-X Series | UH-I Series / AG-I Series | Previous Testing Machines | |
Class 1 |
Use the amp built into the UH-X/FX and AG-X, or an ESA amp.
|
Use an ESA amp.
Precision: Class 1 corresponds to the following measurement ranges.
• DT50-10: 0 mm to 1 mm • DT50-5: 0 mm to 1.25 mm • DT50-2: 0 mm to 1 mm • DT50-50 is not compatible. |
Retrofit the X type controller, and then use the amp built into the UH-X/FX and AG-X, or an ESA amp.
|
Strain control |
Retrofit the X type controller, and then use the amp built into the UH-X/FX and AG-X, or an ESA amp.
|