DUH-210/DUH-210S - Especificações

Dynamic Ultra Micro Hardness Tester

DUH-211

P/N

344-04250-XX (Windows 10)

Required PC Specifications : Expansion Bus

PCI Express ×1 (full-length), 2 slots min. more than one ×1 slot is reguired (or slot size ×2 or more)

Specimen Stage : Vertical Distance

Approx. 60 mm

Specimen Stage : Area

Approx. 125 (W) × 125 (L) mm

Specimen Stage : Stage Movement Range

25 mm in both X and Y directions

Specimen Stage : Specimen Holder

Specimen dimensions (i.e., 8 (thickness) × 30 (width) mm) when thin-type attachment (type 3) is used

Required PC Specifications : OS

Windows® 10 Pro (64 bit edition)

Required PC Specifications : Disk Drives

CD-ROM drive

Utilities : Power Supply

Single phase, AC 100–115 V ± 10%, AC 230 V ± 10% (Ground resistance 100 Ω max.)

Micrometer : Effective Measurement Range

200 µm (with ×50 objective lens)

Utilities : Power Consumption

Approx. 100 W (not including power consumption of PC)

Utilities : Grounding

The ground terminal on 3-prong connectors must be properly grounded with grounding resistance at 100 Ω or less.

Utilities : Temperature

Recommended temperature: 23±1°C,Allowable range: 10°C to 35°C

Utilities : Vibration

Horizontal vibration: 0.017 Gal max. (at 10 Hz or more), 0.01 µm max. (at less than 10 Hz),Vertical vibration: 0.010 Gal max. (at 10 Hz or more), 0.005 µm max. (at less than 10 Hz)

Utilities : Humidity

80% max. (no condensation)

External Dimensions

Tester: Approx. 355 (W) × 405 (D) × 530 (H) mm
Control unit: Approx. 315 (W) × 375 (D) × 110 (H) mm

Micrometer : Minimum Measurement Increment

0.01 µm/pulse

Micrometer : Detector

Optical encoder

Loading Unit : Loading Method

Electromagnetic coil

Displacement
Measurement Unit : Linearity

±2% of full scale (20 µm)

Loading Unit : Test Force Range

Full scale of 0.1 to 1,961 mN

Loading Unit : Test Force Accuracy

±19.6 µN or ±1% of displayed test force, whichever is greater

Loading Unit : Minimum Measurement Increment

0.196 µN (for a test force not exceeding 1.96 mN)

Displacement
Measurement Unit : Measurement Method

Differential transformer

Displacement
Measurement Unit : Measurement Range

0 to 10 µm

Displacement
Measurement Unit : Minimum Measurement Increment

0.0001 µm

Indenter : Type

Triangular pyramid indenter with tip angle of 115° (Vickers indenter and Knoop indenter are available as options.)

Micrometer : Collimation Method

Direct connection between encoder and control handle; synchronized movement of two indexes

Indenter : Tip Radius

0.1 µm max.

Optical Monitor : Total Magnification (microscope)

×500

Optical Monitor : Objective Lens

×50 (Up to 2 lenses can be attached.)

Optical Monitor : Eyepiece

×10

Optical Monitor : Lighting Method

Reflected illumination

Optical Monitor : Light Source (lamp)

LED: 3 W, 3 V

Optical Monitor : Light-Path Switching

Observation or photograph (selectable)

Peso

Tester: Approx. 60 kg
Control unit: Approx. 5 kg

Veja mais
Mostre menos

DUH-211S

P/N

344-04251-XX (Windows 10)

Required PC Specifications : Expansion Bus

PCI Express ×1 (full-length), 2 slots min. more than one ×1 slot is reguired (or slot size ×2 or more)

Specimen Stage : Vertical Distance

Approx. 60 mm

Specimen Stage : Area

Approx. 125 (W) × 125 (L) mm

Specimen Stage : Stage Movement Range

25 mm in both X and Y directions

Specimen Stage : Specimen Holder

Specimen dimensions (i.e., 8 (thickness) × 30 (width) mm) when thin-type attachment (type 3) is used

Required PC Specifications : OS

Windows® 10 Pro (64 bit edition)

Required PC Specifications : Disk Drives

CD-ROM drive

Utilities : Power Supply

Single phase, AC 100–115 V ± 10%, AC 230 V ± 10% (Ground resistance 100 Ω max.)

Micrometer : Effective Measurement Range

200 µm (with ×50 objective lens)

Utilities : Power Consumption

Approx. 100 W (not including power consumption of PC)

Utilities : Grounding

The ground terminal on 3-prong connectors must be properly grounded with grounding resistance at 100 Ω or less.

Utilities : Temperature

Recommended temperature: 23±1°C,Allowable range: 10°C to 35°C

Utilities : Vibration

Horizontal vibration: 0.017 Gal max. (at 10 Hz or more), 0.01 µm max. (at less than 10 Hz),Vertical vibration: 0.010 Gal max. (at 10 Hz or more), 0.005 µm max. (at less than 10 Hz)

Utilities : Humidity

80% max. (no condensation)

External Dimensions

Tester: Approx. 355 (W) × 405 (D) × 530 (H) mm
Control unit: Approx. 315 (W) × 375 (D) × 110 (H) mm

Micrometer : Minimum Measurement Increment

0.01 µm/pulse

Micrometer : Detector

Optical encoder

Loading Unit : Loading Method

Electromagnetic coil

Displacement
Measurement Unit : Linearity

±2% of full scale (20 µm)

Loading Unit : Test Force Range

Full scale of 0.1 to 1,961 mN

Loading Unit : Test Force Accuracy

±19.6 µN or ±1% of displayed test force, whichever is greater

Loading Unit : Minimum Measurement Increment

0.196 µN (for a test force not exceeding 1.96 mN)

Displacement
Measurement Unit : Measurement Method

Differential transformer

Displacement
Measurement Unit : Measurement Range

0 to 10 µm

Displacement
Measurement Unit : Minimum Measurement Increment

0.0001 µm

Indenter : Type

Triangular pyramid indenter with tip angle of 115° (Vickers indenter and Knoop indenter are available as options.)

Micrometer : Collimation Method

Direct connection between encoder and control handle; synchronized movement of two indexes

Indenter : Tip Radius

0.1 µm max.

Optical Monitor : Total Magnification (microscope)

×500

Optical Monitor : Objective Lens

×50 (Up to 2 lenses can be attached.)

Optical Monitor : Eyepiece

×10

Optical Monitor : Lighting Method

Reflected illumination

Optical Monitor : Light Source (lamp)

LED: 3 W, 3 V

Optical Monitor : Light-Path Switching

Observation or photograph (selectable)

Peso

Tester: Approx. 60 kg
Control unit: Approx. 5 kg

Veja mais
Mostre menos

DUH-211

Test force range

0.1 to 1,961 mN (0.01 to 200 gf)

Indentation depth range

0 to 10 µm

Minimum display

0.0001 µm

Testing mode

3 types (211 model), 7 types (211S model)

Veja mais
Mostre menos

DUH-211S

Test force range

0.1 to 1,961 mN (0.01 to 200 gf)

Indentation depth range

0 to 10 µm

Minimum display

0.0001 µm

Testing mode

3 types (211 model), 7 types (211S model)

Veja mais
Mostre menos
{"title":"Downloads das folhas de especifica\u00e7\u00f5es","description":"Baixe as folhas de especifica\u00e7\u00f5es mais recentes.","source":"product","key":3462,"max":30,"filter_types":["spec_sheet"],"link_title":"Ver outros downloads","link_url":""} {"title":"Downloads","description":"Baixe o cat\u00e1logo mais recente","source":"product","key":3462,"max":30,"filter_types":["brochures"],"link_title":"Ver outros downloads","link_url":"","pdf_links":[]}