    USF-2000A - Especificações

    Ultrasonic Fatigue Testing System

    USF-2000A

    Test frequency

    20 kHz ±500 Hz (recommended test range: 20 kHz ±30 Hz)
    Note: The test frequency is determined from the resonance frequency of the sample.

    Vibration at horn tip

    ±10 to ±50 μm

    Test stress range

    Stress given ±10 to ±50 μm displacement of sample
    Note: Stress values depend on sample shape and physical property values.

    Stress ratio

    -1

    Testable materials

    Materials that can be resonated at 20 kHz and generate minimal heat during resonance
    Example: High-strength steel, duralumin, titanium alloy, aluminum, etc.

    Not-testable materials

    - Materials that cannot resonate at 20 kHz
    - Materials for which samples are difficult to attach
    - Materials that generate significant heat during resonance at 20 kHz, due to friction
    Examples: Resins, ceramics, etc.

    Power Requirements

    3-phase 200 V: 2 kVA (air compressor) , 1-phase 200V: 3.5 kVA (ultrasonic fatigue testing system) , 1-phase 100V: 1 kVA (computer, displacement logger, air dryer, etc.)

    Standard contents

    USF-2000A Ultrasonic Fatigue Testing System main unit (including table), ultrasonic resonance system, control computer,ultrasonic testing control and measurement software, and cooling unit (air dryer and compressed air lines)
    Note: Air compressor for cooling is not included.

    Required optional products

    Air compressor (for regions with 50 Hz or 60 Hz power) 3-phase 200 V: 2 kVA
    Displacement measuring system (eddy current displacement gauge with 0.5 μm resolution)
    Note: A high-speed data logger or digital oscilloscope is required separately for reading voltages output from the displacement gauge.
    Displacement gauge calibrator (CDE-25 C1 high-performance micrometer)

    Note: Systems can be selected without an air compressor in cases where the customer will supply the compressed air. A 150 L/m flow rate of compressed air at a minimum 0.2 MPa is required.

    Veja mais
    Mostre menos

    USF-2000A

    Test frequency

    20 kHz ± 500 Hz

    Test stress

    180 to 900 MPa (in the case of a steel circular taper specimen)

    Stress ratio

    -1